Boonton 4530 Peak Power Meter Service Manual Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Boonton 4530 Peak Power Meter Service Manual. Boonton 4530 Peak Power Meter Service Manual User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 37
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY
1
SERVICE AND CALIBRATION MANUAL
BOONTON ELECTRONICS
4530 SERIES
RF PEAK POWER METER
MODELS 4531 AND 4532
REV DATE 9/14/07
P/N 98405100A
BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY
25 EASTMANS ROAD PARSIPPANY, NEW JERSEY 07054-0465
TELEPHONE: (973) 386-9696 FAX: (973) 386-9191 E-MAIL: boonto[email protected]
Vue de la page 0
1 2 3 4 5 6 ... 36 37

Résumé du contenu

Page 1

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 1 SERVICE AND CALIBRATION MANUAL BOONTON ELECTRONICS 4530 SERIES RF PEAK POWER MET

Page 2 - CONTENTS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 10 12) CHANNEL 2 PEAK ACQUISITION TESTS (Model 4532 only) a. Select internal calibrator and

Page 3

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 11 Section 2: 4530 Series Calibration Procedure Note: The following procedures assume that the

Page 4

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 12 1) BASIC SYSTEM CHECK AND SETUP a. Turn ON rear panel power switch b. Press ON/STBY key

Page 5

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 13 2) Calibrator Signal Check a. Press ON/STBY key on front panel. The instrument should pow

Page 6

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 14 3) 50MHz Calibrator RF Alignment Note: The calibrator alignment adjustment is sealed at the

Page 7

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 15 4) CW Channel Input Current Offset Adjustment Warning Extreme care should be used when

Page 8

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 16 500K. e. Connect the 2510 output to the 4530’s Sensor 1 (or 2) input using a CW SENSOR DAT

Page 9

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 17 Construct the calibration setup as shown below. The use of a second 4530 (“reference powerm

Page 10

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 18 [Set Ref] means to place Meas4530 into reference (ratiometric) mode and load the current po

Page 11

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 19 j. [Set Ref]: First [Select 2520], and then once the reading has settled, place Meas4530

Page 12

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 2 CONTENTS Contents 2 Section 1. 4530 Series Test Procedure 3 Sectio

Page 13 - 2) Calibrator Signal Check

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 20 OR Send “CAL1:INT:ZERO” to the Meas4530, and wait for the zeroing process to complete. u.

Page 14

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 21 e. Zero the reference powermeter as follows: Pressing the Zero/CAL key. Select the appropr

Page 15

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 22 OR Send “CALC1:MATH CH1” to the UUT. n. Perform an AutoCal on the sensor using the UUT’s

Page 16

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 23 Section 3: 4530 Series Block Diagrams

Page 17 - Test Setup

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 24

Page 18

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 25

Page 19

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 26

Page 20

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 27 Section 4: 4530 Series Drawings

Page 21

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 28

Page 22

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 29

Page 23

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 3 Section 1: 4530 Series Test Procedure Note: The following procedures assume that the technic

Page 24

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 30

Page 25

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 31

Page 26

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 32 4530 Host Printed Wiring Assembly

Page 27 - Drawings

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 33 Appendix: Entering Data into Boonton Peak and CW Sensors The information supplied in this

Page 28

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 34 Format type, sensor type, month of manufacture, day of manufacture, year of manufacture, se

Page 29

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 35 The following is a properly composed string to be sent to a sensor: RD-S-SLOW 2,0.50,18.00,

Page 30

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 36 SENSOR MESSAGE This is the message that displays on the bottom of the screen wh

Page 31

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 37 normal operation, this feature provides for power-up without repeating the Auto-ca

Page 32

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 4 4) TESTING DSP BOARD POWER SUPPLY a. Locate the DSP Board: large pc board on bottom of chass

Page 33

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 5 k. Press and hold the ON/STBY key. Verify that the instrument turns off (enters standby

Page 34 - NOTE ON CALIBRATION FACTORS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 6 5. Set up the 4400/4500 instrument as the source for the pulsed calibrator signal as follow

Page 35 - ADJUSTING CAL FACTORS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 7 l. Monitor the VIDEO OUT 1 on the rear panel of the 4530 with the oscilloscope. Verify that

Page 36 - CW GAIN COEFFICIENTS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 8 9) CHANNEL 1 CW ACQUISITION TESTS a. Select internal calibrator and turn on at 0dBm: Display

Page 37

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 9 b. Plug in the CW sensor cable into the 4530 Sensor 2 connector and connect a model 51075 CW

Commentaires sur ces manuels

Pas de commentaire