BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY
1
SERVICE AND CALIBRATION MANUAL
BOONTON ELECTRONICS
4530 SERIES
RF PEAK POWER METER
MODELS 4531 AND 4532
REV DATE 9/14/07
P/N 98405100A
BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY
25 EASTMANS ROAD PARSIPPANY, NEW JERSEY 07054-0465
TELEPHONE: (973) 386-9696 FAX: (973) 386-9191 E-MAIL: boonto[email protected]
Vue de la page 0
1 2 3 4 5 6 ... 36 37

Résumé du contenu

Page 1

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 1 SERVICE AND CALIBRATION MANUAL BOONTON ELECTRONICS 4530 SERIES RF PEAK POWER MET

Page 2 - CONTENTS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 10 12) CHANNEL 2 PEAK ACQUISITION TESTS (Model 4532 only) a. Select internal calibrator and

Page 3

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 11 Section 2: 4530 Series Calibration Procedure Note: The following procedures assume that the

Page 4

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 12 1) BASIC SYSTEM CHECK AND SETUP a. Turn ON rear panel power switch b. Press ON/STBY key

Page 5

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 13 2) Calibrator Signal Check a. Press ON/STBY key on front panel. The instrument should pow

Page 6

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 14 3) 50MHz Calibrator RF Alignment Note: The calibrator alignment adjustment is sealed at the

Page 7

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 15 4) CW Channel Input Current Offset Adjustment Warning Extreme care should be used when

Page 8

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 16 500K. e. Connect the 2510 output to the 4530’s Sensor 1 (or 2) input using a CW SENSOR DAT

Page 9

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 17 Construct the calibration setup as shown below. The use of a second 4530 (“reference powerm

Page 10

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 18 [Set Ref] means to place Meas4530 into reference (ratiometric) mode and load the current po

Page 11

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 19 j. [Set Ref]: First [Select 2520], and then once the reading has settled, place Meas4530

Page 12

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 2 CONTENTS Contents 2 Section 1. 4530 Series Test Procedure 3 Sectio

Page 13 - 2) Calibrator Signal Check

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 20 OR Send “CAL1:INT:ZERO” to the Meas4530, and wait for the zeroing process to complete. u.

Page 14

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 21 e. Zero the reference powermeter as follows: Pressing the Zero/CAL key. Select the appropr

Page 15

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 22 OR Send “CALC1:MATH CH1” to the UUT. n. Perform an AutoCal on the sensor using the UUT’s

Page 16

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 23 Section 3: 4530 Series Block Diagrams

Page 17 - Test Setup

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 24

Page 18

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 25

Page 19

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 26

Page 20

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 27 Section 4: 4530 Series Drawings

Page 21

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 28

Page 22

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 29

Page 23

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 3 Section 1: 4530 Series Test Procedure Note: The following procedures assume that the technic

Page 24

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 30

Page 25

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 31

Page 26

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 32 4530 Host Printed Wiring Assembly

Page 27 - Drawings

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 33 Appendix: Entering Data into Boonton Peak and CW Sensors The information supplied in this

Page 28

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 34 Format type, sensor type, month of manufacture, day of manufacture, year of manufacture, se

Page 29

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 35 The following is a properly composed string to be sent to a sensor: RD-S-SLOW 2,0.50,18.00,

Page 30

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 36 SENSOR MESSAGE This is the message that displays on the bottom of the screen wh

Page 31

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 37 normal operation, this feature provides for power-up without repeating the Auto-ca

Page 32

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 4 4) TESTING DSP BOARD POWER SUPPLY a. Locate the DSP Board: large pc board on bottom of chass

Page 33

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 5 k. Press and hold the ON/STBY key. Verify that the instrument turns off (enters standby

Page 34 - NOTE ON CALIBRATION FACTORS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 6 5. Set up the 4400/4500 instrument as the source for the pulsed calibrator signal as follow

Page 35 - ADJUSTING CAL FACTORS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 7 l. Monitor the VIDEO OUT 1 on the rear panel of the 4530 with the oscilloscope. Verify that

Page 36 - CW GAIN COEFFICIENTS

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 8 9) CHANNEL 1 CW ACQUISITION TESTS a. Select internal calibrator and turn on at 0dBm: Display

Page 37

BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY 9 b. Plug in the CW sensor cable into the 4530 Sensor 2 connector and connect a model 51075 CW