Boonton 9240 RF Voltmeter Manuel d'utilisateur Page 67

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 100
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 66
Boonton 9240 Series RF Voltmeter
5.5.5 FETCh Queries
The FETCh? group of queries is used to return specific measurement data from a measurement cycle that has been INITiated
and is complete or free-running. FETCh? performs the data output portion of the measurement. FETCh? does not start a
new measurement, so a series of FETCh? queries may be used to return more than one set of processed measurements from a
complete set of acquired data. FETCh? usually returns the current value of measurements, and should be used anytime free
running data acquisition is taking place (INITiate:CONTinuous ON). If FETCh? is used for single measurements
(INITiate:CONTinuous OFF), no data will be returned until a measurement has been INITiated and is complete.
FETCh:CW:VOLTage?
Description: Return current average amplitude reading in channel units.
Syntax: FETCh[1|2]:CW:VOLTage?
Returns: CC, average voltage
Where CC is the measurement condition code.
FETCh:KEY?
Description: Return the key code of the last key depressed; e.g. MENU = 1.
Syntax: FETCh:KEY?
Returns: key code
Key Code
Menu 1
Sensor 2
FREQ 4
AVG 8
Zero/Cal 16
REF Level 32
Up Arrow 64
Left Arrow 128
Enter 256
Right Arrow 512
Down Arrow 1024
Remote Operation