Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-43
INITiate[:IMMediate[:ALL]]
Description: Starts a single measurement cycle when INITiate:CONTinuous is set to OFF. In Modulated
mode, the measurement will complete once the power has been integrated for the full FILTer time.
In Pulse mode, enough trace sweeps must be triggered to satisfy the AVERaging setting. In
Statistical mode, acquisition stops once the terminal condition(s) are met. In each case, no reading
will be returned until the measurement is complete. This command is not valid when
INITiate:CONTinuous is ON, however, by convention this situation does not result in a SCPI
error.
Syntax: INITiate[:IMMediate[:ALL]]
Argument: None
Valid Modes: Any
Restrictions: INITiate:CONTinuous must be OFF