Boonton 4500B Peak Power Meter Manuel d'utilisateur Page 355

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 372
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 354
Boonton 4500B RF Peak Power Analyzer
Maintenance
7-23
Table 7-14. 56340/540, 57340/540, 59340
Sensor Frequency Calfactor Accuracy
Measured
Frequency Reference Minimum CH1 CH2 Maximum
(GHz) (dBm) (dB) (dB) (dB) (dB)
0.10 ________ -0.14 ________ ________ 0.14
0.50 ________ -0.14 ________ ________ 0.14
2.00 ________ -0.15 ________ ________ 0.15
4.00 ________ -0.19 ________ ________ 0.19
6.00 ________ -0.19 ________ ________ 0.19
8.00 ________ -0.20 ________ ________ 0.20
10.00 ________ -0.20 ________ ________ 0.20
12.00 ________ -0.20 ________ ________ 0.20
14.00 ________ -0.21 ________ ________ 0.21
16.00 ________ -0.21 ________ ________ 0.21
18.00 ________ -0.21 ________ ________ 0.21
20.00 ________ -0.26 ________ ________ 0.26
22.00 ________ -0.26 ________ ________ 0.26
24.00 ________ -0.26 ________ ________ 0.26
26.50 ________ -0.26 ________ ________ 0.26
28.00 ________ -0.31 ________ ________ 0.31
32.00 ________ -0.31 ________ ________ 0.31
36.00 ________ -0.31 ________ ________ 0.31
40.00 ________ -0.31 ________ ________ 0.31
Vue de la page 354
1 2 ... 350 351 352 353 354 355 356 357 358 359 360 ... 371 372

Commentaires sur ces manuels

Pas de commentaire