Boonton PIM 31 User Manual Manuel d'utilisateur Page 87

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 110
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 86
87
Connect the Coupler, Low PIM Load and Power Meter to the PIM 31 test
system (use torque wrench)
Set the PIM 31 to Analyzer Mode
Set frequencies of carrier signals 1 and 2 to the start and end frequency of
the provided frequency band.
Set the power level of each signal to 40 dBm (20W)with ALC: ON
Switch RF Power ON
With offset compensation of the coupler (30dB) the power meter should
read 43dBm +/- 1.0dB
If the reading deviates, switch RF power OFF, disconnect connections,
reconnect and repeat the test.
.
If the reading is off by more than 1 dBm, the deviation could be caused
by an malfunctioning attenuator or amplifier in the PIM 31 Test System.
In this case, contact your nearest Boonton Service Center
.
With the reading within limits, continue procedure:
Increase Signal Levels by 1dB,
Measure power
Repeat until Signal Levels are both at 43dBm.
With offset compensation of the coupler (30dB) the
power meter should
read now: 46dBm +/- 1.0dB
Switch “RF Power OFF
Vue de la page 86
1 2 ... 82 83 84 85 86 87 88 89 90 91 92 ... 109 110

Commentaires sur ces manuels

Pas de commentaire