Boonton 4540 Series RF Power Meter
Remote Operation
5.6.9 MEMory Subsystem
The MEMory group of commands is used to save and recall instrument operating configurations, and to edit and review user-
supplied frequency dependent offset (FDOF) tables for external devices in the signal path. Up to four configurations may be
saved, and two frequency dependent offset tables. Note, however that assigning a stored FDOF table to a particular
measurement channel is not a MEMory command; it is handled through the SENSe subsystem.
MEMory:SNSR:CFFAST?
Description: Return the sensor high bandwidth (FAST) frequency cal-factor table.
Syntax: MEMory:SNSR[1|2]:CFFAST?
Argument: None, query only.
Valid Modes: Any
MEMory:SNSR:CFSLOW?
Description: Return the sensor low bandwidth (SLOW) frequency cal-factor table.
Syntax: MEMory:SNSR[1|2]:CFSLOW?
Argument: None, query only.
Valid Modes: Any
MEMory:SNSR:CWRG?
Description: Return sensor AC cal data.
Syntax: MEMory:SNSR[1|2]:CWRG?
Argument: None, query only.
Valid Modes: Any
MEMory:SNSR:CWSH?
Description: Return sensor smart shaping table.
Syntax: MEMory:SNSR[1|2]:CWSH?
Argument: None, query only.
Valid Modes: Any