Boonton 4540 Peak Power Meter User Manual Manuel d'utilisateur Page 183

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 270
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 182
Boonton 4540 Series RF Power Meter
Remote Operation
5-49
READ:MARKer:AVERage?
Description: For the specified channel and marker, return the average power or voltage at the marker. The
units are the same as the specified channel.
Syntax: READ[1|2]:MARKer[1|2]:AVERage?
Returns: CC, average power or voltage at marker
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:MARKer:DELTa?
Description: For the specified channel return the difference between MK1 and MK2. The units will be the
same as marker units.
Syntax: READ[1|2]:MARKer:DELTa?
Returns: CC, (MK1 MK2)
Where CC is the measurement condition code.
Valid Modes: All
READ:MARKer:MAXimum?
Description: For the specified channel and marker, return the maximum power or voltage at the marker. The
units are the same as the specified channel.
Syntax: READ[1|2]:MARKer[1|2]:MAXimum?
Returns: CC, maximum power or voltage at marker
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated.
READ:MARKer:MINimum?
Description: For the specified channel and marker, return the minimum power or voltage at the marker. The
units will be the same as the specified channel.
Syntax: READ[1|2]:MARKer[1|2]:MINimum?
Returns: CC, minimum power or voltage at marker
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated.
Vue de la page 182
1 2 ... 178 179 180 181 182 183 184 185 186 187 188 ... 269 270

Commentaires sur ces manuels

Pas de commentaire