Boonton 4540 Series RF Power Meter
Remote Operation
5-48
READ:INTERval:MINFilt?
Description: For the specified channel, return the minimum filtered power or voltage in the time interval
between marker 1 and marker 2. The units will be the same as the specified channel.
Syntax: READ[1|2]:INTERval:MINFilt?
Returns: CC, minimum filtered power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:INTERval:MAXimum?
Description: For the specified channel, return the maximum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax: READ[1|2]:INTERval:AVERage?
Returns: CC, maximum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:INTERval:MINimum?
Description: For the specified channel, return the minimum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.
Syntax: READ[1|2]:INTERval:MINimum?
Returns: CC, minimum instantaneous power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:INTERval:PKAVG?
Description: For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.
Syntax: READ[1|2]:INTERval:PKAVG?
Returns: CC, peak-to-average ratio of power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated