Boonton 4540 Series RF Power Meter
Operation
4-46
Main > Time > Trig Delay
Description: Set the trigger delay time in seconds with respect to the trigger. Positive values cause the
actual trigger to occur after the trigger condition is met. This places the trigger event to
the left of the trigger point on the display, and is useful for viewing events during a pulse,
some fixed delay time after the rising edge trigger. Negative trigger delay places the
trigger event to the right of the trigger point on the display, and is useful for looking at
events before the trigger edge.
Menu Type: Numeric Value.
Range: Timebase setting Trigger Delay range
5 ns/div to 500 ns/div 4 ms to +100 ms
1 µs/div to 10 ms/div ±4000 divisions
20 ms/div to 3600 sec/div 40 to +100 sec
SCPI Command: TRIGger:DELay <numeric_value>
Main > Time > Position Cntrl
Description: Selects the trigger position mode. Preset provides three trigger positions, while Vernier
allows the user to select any positon from -30.0 to + 30.0 divisions in 0.1 division
increments relative to the trigger event.
Menu Type: Toggle.
Choices: Vernier, Preset
SCPI Command: None. Note: Selecting TRIGger:VERNier or TRIGger:POSition forces the position
cntrl setting.