Boonton 4540 Series RF Power Meter
Remote Operation
5-69
STATus:QUEStionable:CALibration:NTRansition
Description: Set or return the value of the negative transition filter bit mask for the Questionable Calibration
status group. Setting a bit in the positive transition filter causes a 1 to 0 (negative) transition in the
corresponding bit of the Questionable Calibration Condition register to cause a 1 to be written in
the associated bit of the Questionable Calibration Event register.
Syntax: STATus:QUEStionable:CALibration:NTRansition <numeric_value>
Argument: <numeric_value> = 0 to 32767
Returns: 16-bit register value (0 to 32767)
Valid Modes: Any
STATus:QUEStionable:CALibration:PTRansition
Description: Set or return the value of the positive transition filter bit mask for the Questionable Calibration
status group. Setting a bit in the positive transition filter causes a 0 to 1 (positive) transition in the
corresponding bit of the Questionable Calibration Condition register to cause a 1 to be written in
the associated bit of the Questionable Calibration Event register.
Syntax: STATus:QUEStionable:CALibration:PTRansition <numeric_value>
Argument: <numeric_value> = 0 to 32767
Returns: 16-bit register value (0 to 32767)
Valid Modes: Any