Boonton 4540 Series RF Power Meter
Remote Operation
5-40
OUTPut:EXTernal:PULse:PERWID
Description: Set or return the pulse period <arg1> and pulse width <arg2> in seconds in the Variable pulse
control mode of the external 1 GHz calibrator. This parameter has no effect in the preset pulse
control mode or the external pulse source mode.
Syntax: OUTPut:EXTernal:PULse:PERWID <arg1> <arg2>
Argument: <arg1> = 28e-6 to 131.07e-3 (Pulse period)
<arg2> = 7e-6 to 65.535e-3 (Pulse width)
Valid Modes: Any
OUTPut:EXTernal:PULse:SOURce
Description: Set or return the pulse modulation source for the external 1 GHz calibrator. If set to INT, the
internal pulse generator is used. EXT selects the signal from the calibrator’s rear-panel EXT
PULSE input. Note this input requires TTL compatible signal levels.
Syntax: OUTPut:EXTernal:PULse:SOURce <character data>
Argument: <character data> = {INT, EXT}
Valid Modes: Any
OUTPut:EXTernal:SIGNal
Description: Set or return the on/off state of the external 1 GHz calibrator output signal.
Syntax: OUTPut:EXTernal:SIGNal <Boolean>
Argument: <Boolean> = { 0, 1, OFF, ON }
Valid Modes: Any
OUTPut:INTernal:LEVel[:POWer]
Description: Set or return the power level of the internal 50 MHz calibrator output signal.
Syntax: OUTPut:INTernal:LEVel:POWer < numeric_value >
Argument: <character data> = -60.0 to +20.0 dBm (0.1dB resolution)
Valid Modes: Any